csirt@kutim:~/threat-intel/cve/CVE-2026-17583

Katalog CVE

CVSS 8.4 · High EPSS 0,14%

Kerentanan Missing Support for Integrity Check pada thermofisher abi prism 310 data collection software, thermofisher abi prism 3100/3100-avant data collection software +6 lainnya. Skor CVSS 3.1 8.4 (High) menurut hq.dhs.gov.

Deskripsi (NVD)

The affected Thermo Fisher Applied Biosystems Genetic Analyzers are vulnerable because .fsa/.hid output files can be edited. An attacker could tamper with these files, altering DNA data and resulting in inaccurate DNA test outcomes.

Rincian

Terbit di NVD
5 Agustus 2026
Terakhir diubah
26 Agustus 2026
Status NVD
Analyzed
CVSS 3.1
8.4 (High), sumber hq.dhs.gov
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H · uraikan di kalkulator
Jenis kelemahan (CWE)
CWE-353: Missing Support for Integrity Check
EPSS (30 hari)
0,14% peluang dieksploitasi; lebih tinggi dari 2,9% CVE lain

Produk dan versi terdampak

VendorProdukVersi terdampak
thermofisher abi prism 310 data collection software Semua versi hingga 3.1
thermofisher abi prism 3100/3100-avant data collection software Semua versi hingga 2.0
thermofisher applied biosystems 3130 series data collection software Semua versi hingga 4.1
thermofisher applied biosystems 3500/3500xl series data collection software Semua versi sebelum 4.0.3
thermofisher applied biosystems 3730/3730xl series data collection software Semua versi sebelum 5.0.3
thermofisher applied biosystems seqstudio flex series instrument software Semua versi sebelum 1.2.1
thermofisher applied biosystems seqstudio genetic analyzer data collection software Semua versi sebelum 1.2.6
thermofisher genemapper id-x Semua versi sebelum 1.7.4

> Data dari konfigurasi CPE di NVD.

Referensi dan perbaikan

Sumber: NVD · CVE.org · EPSS. Halaman ini disusun otomatis dari data tersebut; pembaruan terakhir 5 Okt 2026, 20:11 WITA. Kalau sistem di Perangkat Daerah memakai produk ini dan butuh bantuan, laporkan lewat kanal pelaporan.