csirt@kutim:~/threat-intel/cve/CVE-2025-54304

Katalog CVE

CVSS 9.8 · Critical EPSS 0,50%

Kerentanan Exposure of Sensitive Information to an Unauthorized Actor pada thermofisher ion torrent onetouch 2 firmware. Skor CVSS 3.1 9.8 (Critical) menurut CISA ADP.

Deskripsi (NVD)

An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. When they are powered on, an X11 display server is started. The display server listens on all network interfaces and is accessible over port 6000. The X11 access control list, by default, allows connections from 127.0.0.1 and 192.168.2.15. If a device is powered on and later connected to a network with DHCP, the device may not be assigned the 192.168.2.15 IP address, leaving the display server accessible by other devices on the network. The exposed X11 display server can then be used to gain root privileges and the ability to execute code remotely by interacting with matchbox-desktop and spawning a terminal. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.

Rincian

Terbit di NVD
4 Desember 2025
Terakhir diubah
17 Juni 2026
Status NVD
Analyzed
CVSS 3.1
9.8 (Critical), sumber CISA ADP
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H · uraikan di kalkulator
Jenis kelemahan (CWE)
CWE-200: Exposure of Sensitive Information to an Unauthorized Actor
EPSS (30 hari)
0,50% peluang dieksploitasi; lebih tinggi dari 41,0% CVE lain

Produk dan versi terdampak

VendorProdukVersi terdampak
thermofisher ion torrent onetouch 2 firmware Semua versi

> Data dari konfigurasi CPE di NVD.