csirt@kutim:~/threat-intel/cve/CVE-2025-15653

Katalog CVE

CVSS 6.8 · Medium EPSS 0,17%

Kerentanan Exposure of Resource to Wrong Sphere. Skor CVSS 3.1 6.8 (Medium) menurut vulncheck.com.

Deskripsi (NVD)

Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.

Rincian

Terbit di NVD
2 Juni 2026
Terakhir diubah
22 Juli 2026
Status NVD
Awaiting Analysis
CVSS 3.1
6.8 (Medium), sumber vulncheck.com
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H · uraikan di kalkulator
Jenis kelemahan (CWE)
CWE-668: Exposure of Resource to Wrong Sphere
EPSS (30 hari)
0,17% peluang dieksploitasi; lebih tinggi dari 5,8% CVE lain

Produk dan versi terdampak

NVD belum mencantumkan data produk (CPE) untuk CVE ini. Produk yang terdampak biasanya disebut di deskripsi dan di advisory vendor pada daftar referensi.

Referensi dan perbaikan