csirt@kutim:~/threat-intel/cve/CVE-2025-14095

Katalog CVE

CVSS 6.8 · Medium EPSS 0,17%

Kerentanan Improper Access Control. Skor CVSS 3.1 6.8 (Medium) menurut CNA.

Deskripsi (NVD)

A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software.  Other related CVE's are CVE-2025-14096 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required configuration for Exposure: Physical access to the analyzer is needed. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication.                                                                                                                                                                                        Note: CVSS score 6.8 when underlying OS is Windows 7 or Windows XP Operating systems and CVSS score 5.7 when underlying OS is Windows 8 or Windows 10 operating systems.

Rincian

Terbit di NVD
17 Desember 2025
Terakhir diubah
7 Oktober 2026
Status NVD
Deferred
CVSS 3.1
6.8 (Medium), sumber CNA
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H · uraikan di kalkulator
Jenis kelemahan (CWE)
CWE-284: Improper Access Control
CWE-693: Protection Mechanism Failure
EPSS (30 hari)
0,17% peluang dieksploitasi; lebih tinggi dari 5,4% CVE lain

Produk dan versi terdampak

NVD belum mencantumkan data produk (CPE) untuk CVE ini. Produk yang terdampak biasanya disebut di deskripsi dan di advisory vendor pada daftar referensi.

Referensi dan perbaikan

Sumber: NVD · CVE.org · EPSS. Halaman ini disusun otomatis dari data tersebut; pembaruan terakhir 8 Okt 2026, 00:25 WITA. Kalau sistem di Perangkat Daerah memakai produk ini dan butuh bantuan, laporkan lewat kanal pelaporan.