Kerentanan Exposure of Private Personal Information to an Unauthorized Actor. Skor CVSS 3.1 5.9 (Medium) menurut hq.dhs.gov.

Deskripsi (NVD)

In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.

Rincian

Terbit di NVD
30 Januari 2025
Terakhir diubah
17 Juni 2026
Status NVD
Deferred
CVSS 3.1
5.9 (Medium), sumber hq.dhs.gov
CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N · uraikan di kalkulator
Jenis kelemahan (CWE)
CWE-359: Exposure of Private Personal Information to an Unauthorized Actor
EPSS (30 hari)
0,79% peluang dieksploitasi; lebih tinggi dari 55,1% CVE lain

Produk dan versi terdampak

NVD belum mencantumkan data produk (CPE) untuk CVE ini. Produk yang terdampak biasanya disebut di deskripsi dan di advisory vendor pada daftar referensi.

Referensi dan perbaikan