CVSS 9.3 · Critical EPSS 0,18%
Kerentanan Improper Restriction of Excessive Authentication Attempts. Skor CVSS 3.1 9.3 (Critical) menurut baxter.com.
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H