CVSS 9.3 · Critical EPSS 0,22%
Kerentanan Improper Physical Access Control. Skor CVSS 3.1 9.3 (Critical) menurut baxter.com.
The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
· uraikan di kalkulator NVD belum mencantumkan data produk (CPE) untuk CVE ini. Produk yang terdampak biasanya disebut di deskripsi dan di advisory vendor pada daftar referensi.