csirt@kutim:~/threat-intel/cve/CVE-2024-48966

Katalog CVE

CVSS 10.0 · Critical EPSS 0,70%

Kerentanan Missing Authentication for Critical Function. Skor CVSS 3.1 10.0 (Critical) menurut baxter.com.

Deskripsi (NVD)

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Rincian

Terbit di NVD
14 November 2024
Terakhir diubah
17 Juni 2026
Status NVD
Deferred
CVSS 3.1
10.0 (Critical), sumber baxter.com
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H · uraikan di kalkulator
Jenis kelemahan (CWE)
CWE-306: Missing Authentication for Critical Function
EPSS (30 hari)
0,70% peluang dieksploitasi; lebih tinggi dari 51,7% CVE lain

Produk dan versi terdampak

NVD belum mencantumkan data produk (CPE) untuk CVE ini. Produk yang terdampak biasanya disebut di deskripsi dan di advisory vendor pada daftar referensi.

Referensi dan perbaikan