ge vivid e95 firmware, ge vivid e90 firmware, ge vivid s70n firmware +13 lainnya
CVSS 6.8 · Medium EPSS 0,43%
Kerentanan Protection Mechanism Failure pada ge vivid e95 firmware, ge vivid e90 firmware +14 lainnya. Skor CVSS 3.1 6.8 (Medium) menurut NVD.
A restricted desktop environment escape vulnerability exists in the Kiosk Mode functionality of affected devices. Specially crafted inputs can allow the user to escape the restricted environment, resulting in access to the underlying operating system. Affected devices include the following GE Ultrasound Products: Vivid products - all versions; LOGIQ - all versions not including LOGIQ 100 Pro; Voluson - all versions; Versana Essential - all versions; Invenia ABUS Scan station - all versions; Venue - all versions not including Venue 40 R1-3 and Venue 50 R4-5
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
· uraikan di kalkulator | Vendor | Produk | Versi terdampak |
|---|---|---|
| ge | vivid e95 firmware | Semua versi |
| ge | vivid e90 firmware | Semua versi |
| ge | vivid s70n firmware | Semua versi |
| ge | vivid t8 firmware | Semua versi |
| ge | vivid t9 firmware | Semua versi |
| ge | vivid iq firmware | Semua versi |
| ge | logiq e10 firmware | Semua versi |
| ge | logiq e9 firmware | Semua versi |
| ge | logiq s8 firmware | Semua versi |
| ge | logiq s7 firmware | Semua versi |
| ge | logiq p9 firmware | Semua versi |
| ge | logiq e9 with xdclear firmware | Semua versi |
| ge | voluson firmware | Semua versi |
| ge | versana essential firmware | Semua versi |
| ge | invenia abus scan station firmware | Semua versi |
| ge | venue go firmware | Semua versi |