csirt@kutim:~/threat-intel/cve/CVE-2020-12048

Katalog CVE

CVSS 7.5 · High EPSS 0,45%

Kerentanan Cleartext Transmission of Sensitive Information pada baxter phoenix x36 firmware. Skor CVSS 3.1 7.5 (High) menurut NVD.

Deskripsi (NVD)

Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.

Rincian

Terbit di NVD
29 Juni 2020
Terakhir diubah
17 Juni 2026
Status NVD
Modified
CVSS 3.1
7.5 (High), sumber NVD
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N · uraikan di kalkulator
Jenis kelemahan (CWE)
CWE-319: Cleartext Transmission of Sensitive Information
EPSS (30 hari)
0,45% peluang dieksploitasi; lebih tinggi dari 37,5% CVE lain

Produk dan versi terdampak

VendorProdukVersi terdampak
baxter phoenix x36 firmware Versi 3.36
baxter phoenix x36 firmware Versi 3.40