csirt@kutim:~/threat-intel/cve/CVE-2020-12040

Katalog CVE

CVSS 9.8 · Critical EPSS 0,95%

Kerentanan Cleartext Transmission of Sensitive Information pada baxter sigma spectrum infusion system firmware. Skor CVSS 3.1 9.8 (Critical) menurut NVD.

Deskripsi (NVD)

Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.

Rincian

Terbit di NVD
29 Juni 2020
Terakhir diubah
17 Juni 2026
Status NVD
Modified
CVSS 3.1
9.8 (Critical), sumber NVD
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H · uraikan di kalkulator
Jenis kelemahan (CWE)
CWE-319: Cleartext Transmission of Sensitive Information
EPSS (30 hari)
0,95% peluang dieksploitasi; lebih tinggi dari 60,1% CVE lain

Produk dan versi terdampak

VendorProdukVersi terdampak
baxter sigma spectrum infusion system firmware Versi 6.0 hingga 6.05
baxter sigma spectrum infusion system firmware Versi 8.0

> Data dari konfigurasi CPE di NVD.

Referensi dan perbaikan