CVSS 7.8 · High EPSS 0,12%
Kerentanan User Interface (UI) Misrepresentation of Critical Information pada draeger infinity explorer c700 firmware. Skor CVSS 3.1 7.8 (High) menurut NVD.
Dräger Infinity Explorer C700 contains a privilege escalation vulnerability that allows attackers to break out of kiosk mode and access the underlying operating system through a specific dialog interaction. Attackers can exploit this kiosk escape to take control of the operating system and cause the device to display incorrect or no information from the connected Delta Family patient monitor.
CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H
· uraikan di kalkulator | Vendor | Produk | Versi terdampak |
|---|---|---|
| draeger | infinity explorer c700 firmware | Semua versi |